To begin a leak test, select the option on the menu page

2
00:00:11.020 --> 00:00:16.230
Set the pressure and test duration then press begin purge.

3
00:00:19.022 --> 00:00:27.050
This will start the pre-programmed leak test. Initially the system will prepare for a leak test.

4
00:00:28.030 --> 00:00:35.040
During the leak test, the glove box will aim to reach this overpressure in the internal chamber, then hold it for the test duration.

5
00:00:35.040 --> 00:00:39.014
Once the test has begun, a countdown will begin.

6
00:00:40.000 --> 00:00:44.011
When the test is finished, a leak rate will be displayed.

7
00:00:44.012 --> 00:00:50.021
Anything under 0.05 is acceptable to maintain an inert environment.

8
00:00:51.000 --> 00:01:02.070
If there is a leak in the glove box, set the glovebox to an overpressure of 5mBar. Leak spray (or soapy water) can be used to identify the leak point.

9
00:01:05.050 --> 00:01:11.190
As the nitrogen is driven out of the leak, this will cause the leak spray to form bubbles.